Wastewater Dilution Test — 50056
Enter dilution annotation windows below. Sensor data and regression update automatically.
Experiment Overview
On 20 May 2026, Lume sensor 50056 was tested in a controlled wastewater dilution series at eight concentration steps: 0%, 1%, 5%, 10%, 25%, 50%, 75%, and 100% raw wastewater in tap water. Each window lasted approximately 8–13 minutes. Two reference sensors were run in parallel: an Aqualog benchtop EEM fluorometer (IFE-corrected, treated as ground truth) and a Turner C3 in-situ fluorometer measuring CDOM (365 nm) and TLF (280 nm). Lab turbidity (NTU) was measured by grab sample.
The LUME sensor swept LED power (64–1024) and SiPM bias (2800–3400 V) continuously throughout the experiment, generating 108,173 raw readings across 44 LED×bias combinations. The calibration combo (LED 512 / bias 3000) was used for fluorescence analysis; TOF backscatter (SPS, kcps/SPAD) provided an independent turbidity channel.
(0–100% WW)
across 44 combos
(Aqualog, Turner C3, NTU)
TLF + low-LED + SPS)
LED/Bias Combo Comparison
Inner filter effect (IFE): At high concentrations, the sample absorbs its own excitation and emitted light, causing the fluorescence signal to plateau then decrease. The IFE onset is a property of the sample's optical density, not the detector — changing LED power scales all signal values proportionally but does not shift the concentration at which the peak occurs (the peak is at OD ≈ 0.05 regardless of I₀). Implication for this dataset: if 50% and 75% are already indistinguishable at LED 512 (flat top of the IFE curve), they will remain indistinguishable at any other LED power because the curve shape is unchanged — only its amplitude shifts. What the comparison below can reveal is whether any combo suffers from additional SiPM microcell saturation on top of IFE (signal plateaus earlier than expected), vs. combos where the SiPM is in its linear range and the true IFE shape is visible. Lines marked ⚠ are non-monotonic (IFE-dominated). Use Normalize to compare curve shapes independent of LED power. One combo per LED level shown (bias nearest to cal).
TLF Signal / Turbidity / Temperature — Sensor 50056
Turner C3 — In-Situ CDOM, TLF & Temperature
Turner C3 grab-sample time series collected in parallel with the dilution experiment. Timestamps are on the Turner’s own clock (not aligned with LUME UTC). Each concentration window has 180 readings at ~1 Hz; background shading shows the annotated windows. Note CDOM continues rising through 75–100% while TLF (tryptophan-like) peaks at ~25% and decreases at high concentrations — inner filter effect on the 280 nm excitation channel.
Wastewater Concentration vs Fluorescence Signal
Lab Turbidity vs TOF Backscatter
Low-end sensitivity (0%→1% step): LUME TLF signal increases by +141.6 counts (+36.8%) with a signal-to-noise ratio of 4.73 — compared to Turner C3 TLF which shows a similar +41.7% relative change but an SNR of only 2.70. LUME resolves the 0%→1% step ~75% more reliably than the Turner TLF on a per-reading basis (180 readings averaged here vs LUME’s continuous stream). Turner CDOM achieves SNR = 5.23 for the same step but measures organic carbon rather than tryptophan-like fluorescence.
Multivariate Regression — Predicted vs Actual Concentration (LOO-CV)
OLS fits using leave-one-out cross-validation: each point is predicted by a model trained on all other windows. R²LOO is an honest out-of-sample estimate — lower than in-sample R² (shown in parentheses). Left: TLF + TOF + temperature. Right: TLF + TOF only (temperature dropped — only 1.4 °C variation, costs a free parameter).
TLF + TOF + Temperature
TLF + TOF only
Multi-Channel Model — Full-Range LOO-CV
The cal combo (LED 512 / bias 3000) saturates above ~25% — mon2 is flat at 50%, 75%, and 100% (SiPM hardware ceiling). At LED 64 / bias 3425 the SiPM stays below saturation, so the inner filter effect is visible: signal peaks at ~50% then decreases at 75% and 100% (SNR ≈ 7.6 between the three). Combined with SPS (TOF turbidity), this 3-predictor OLS achieves R²LOO = 0.999 across all 8 windows. Left: normalized (0–1) channel responses — shows where each channel saturates. Right: LOO-CV parity plot for the combined model.
Normalized Channel Responses
3-Predictor LOO-CV (cal TLF + low-LED TLF + SPS)
Reference Sensor Comparison — Normalized Signal vs Concentration
All signals normalized 0–1 for shape comparison. Aqualog (benchtop EEM, IFE-corrected): both TLF and CDOM are linear to 100%. Turner C3: CDOM (365 nm) nearly linear; TLF (280 nm) peaks near 25% and reverses. LUME in-situ TLF saturates above ~25–50% — same IFE behavior as all in-situ fluorescence sondes.
Sensor Performance vs Aqualog Truth
Each sensor signal (normalized 0–1) vs Aqualog TLF (normalized, ground truth). A perfect sensor lies on the 1:1 line with R² = 1.0. R² shown for full range, linear regime (0–25%), and IFE regime (50–100%).
50–100% wastewater (IFE regime): LUME TLF R² = 0.798 vs Turner C3 TLF R² = 0.289. Turner TLF signal fully reverses direction past ~25% (IFE inverts the response), destroying predictive power. LUME TLF also drops at 100% due to IFE, but attenuates more gradually — signal compression rather than reversal — resulting in nearly 3× higher R² in the high-concentration regime.
Add Annotation
Concentration Annotations
No annotations yet. Add one using the form above.